Operator's Manual
WP700Zi-OM-E-RevA
174
4.
If you want to change the parameter listed, or a measurement characteristic, touch the parameter button
(
P1
for example) alongside the check box. A pop-up menu of parameters categorized by type
appears.
To display parameter icons only, touch the icon button
at the bottom of the menu.
To display the icons in list form, along with an explanation of each parameter, touch the list button
.
Use the
Up/Down
buttons to scroll through the list of icons.
4.
When you make a selection from the parameter icon menu, the setup dialogs for that parameter appear.
You can then change the waveform source and other conditions of the parameter.
5.
If you are setting up an "@level" parameter, make selections for
Level type
(percent or absolute),
Slope
(positive, negative, both), and
Hysteresis
level.
6. Touch
the
Gate
tab, and set the position of the gate posts.
From a Vertical Setup Dialog
1. Touch
Vertical Channelx Setup
on the menu bar.
2. In
the
Cx Vertical Adjust
dialog, touch the Measure button
.
3.
Select a parameter from the pop-up menu. (The
Actions for trace
source defaults to the channel or trace
whose dialog is open. If a parameter, it goes into the next "available" parameter, or the last one if all are
used.)
4.
Select another parameter or touch
Close
.
From a Math Setup Dialog
1. Touch
Math Fx Setup
on the menu bar.
2. In
the
Fx
dialog, touch the Measure button
.
3.
Select a parameter from the pop-up menu. (The
Actions for trace
source defaults to the channel or trace
whose dialog is open. If a parameter, it goes into the next "available" parameter, or the last one if all are
used.)
4.
Select another parameter or touch
Close
.
List of Parameters
Instrument parameters are described as follows. Parameter availability varies based on the options installed on
the oscilloscope (specific references provided in corresponding Notes).
Parameter
Description
Definition
Notes
100BT Fall
(fall)
Fall time between 2 levels
(upper-base, base-lower) of
a 3-level signal (100BT)
Available with ENET option.
100BT Rise
(rise)
Rise time between 2 levels
(Lower-base, base-upper) of
a 3-level signal (100BT)
Available with ENET option.
100BT TIE
(TIE)
Difference between the
measured and ideal times at
level between base and
upper or lower levels of
100BT signal.
Available with ENET option.
100BT Tj
Total jitter from a TIE at level
Available with ENET option.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...