WavePro 7Zi
141
WP700Zi-OM-E-RevA
Characteristic
Trend
Track
Ensures No Lost
Measurement Data
No. Since data can be accumulated over many
acquisitions, and since the oscilloscope takes
time to calculate measurement values and to
display data before the trigger is re-armed, data
can be missed.
Yes. Maximum time period
that can be captured is
limited by acquisition
memory and sampling rate.
In general, Track will be the tool to use if you want to capture a continuous stream of data that is spaced closely
together. Trend can be used if your data is spaced widely apart and longer than the “dead-time” of the
oscilloscope between acquisitions.
Resampling to Deskew
Deskew whenever you need to compensate for different lengths of cables, probes, or anything else that might
cause timing mismatches between signals. Resample a signal on one channel and adjust it in time relative to a
signal on another channel.
Deskewing
1. Touch
Math Math Setup
on the menu bar.
2.
Touch a math function trace tab
F1
through
Fx
[The number of math traces available depends on the
software options loaded on your scope. See Specifications.]
3. Touch
the
single
function button.
4. Touch
inside
the
Source1
field and select a source: channel, math trace, memory location.
5. Touch
inside
the
Operator1
field and select
Deskew
from the
Functions
category.
6.
In the dialog on the right, touch the
Deskew
tab.
7. Touch
inside
the
Delay by
data entry field and type in a time value, using the pop-up keypad.
Rescaling and Assigning Units
This feature allows you to apply a multiplication factor (
a
) and additive constant (
b
) to your waveform:
a
X +
b
. You
can do it in the unit of your choice, depending on the type of application.
Allowable unit
abbreviations
and
descriptions
are as follows:
Abbreviations and Descriptions
1.
(blank)
- No
units
2.
A
- Ampere
3.
C
- Coulomb
4.
CYCLE
-
Cycles
5.
DB
- Decibel
6.
DBC
- Decibel
referred to
carrier
7.
DBM
- Decibel
Milliwatt
8.
DBV
- Decibel
Volts
9.
DBUZ
-
Decibel
Microamp
10.
DEC
- Decade
11.
DIV
-
Divisions
12.
Event
-
Events
13.
F
- Farad
14.
G
- Gram
15.
H
- Henry
16.
HZ
- Hertz
17.
J
- Joule
18.
K
- Degree
Kelvin
19.
CEL
- Degree
Celsius
20.
FAR
- Degree
Farenheit
21.
L
- Liter
22.
M
- Meter
23.
FT
- Foot
24.
IN
- Inch
25.
YARD
- Yard
26.
MILE
- Mile
27.
N
- Newton
28.
OHM
- Ohm
29.
PAL
- Pascal
30.
PCT
- Percent
31.
POISE
- Poise
32.
PPM
- Parts per
million
33.
RAD
- Radian
34.
DEG
- Degree (of
arc)
35.
MNT
- Minute (of
arc)
36.
SAMPLE
-
Sample
37.
SWEEP
- Sweeps
38.
SEC
- Second (of
arc)
39.
S
- Second
40.
SIE
- Siemens
41.
T
- Tesla
42.
UI
- Unit
interval
43.
V
- Volt
44.
VA
- Volt
amps
45.
W
- Watt
46.
WB
- Weber
47.
MIN
- Min
48.
HOUR
- Hour
49.
DAY
- Day
50.
WEEK
-
Week
You can also enter combinations of the above units following the SI rules:
x
for the quotient of two units, the character
/
should be used
x
for the product of two units, the character
.
should be used
x
exponents can be represented by a digit appended to the unit without a space
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...