Operator's Manual
WP700Zi-OM-E-RevA
386
Choosing a Timing Parameter
This table lists the Jitter and Timing Analysis (JTA) parameters and the tasks that they can perform. Additional
analysis and processing of the waveform can be carried out by activating Statistics and using histogram
parameters. For some parameters, one of the variants of JitterTrack can perform the same task.
If You Want To
Use This Timing
Parameter
For Further Processing
Or JitterTrack
measure accuracy of clock, period
or frequency
p@lv
freq@lv
Statistics On
or use
Histogram
Period Jitter
Frequency Jitter
measure pulse width accuracy
wid@lv
Statistics On
or use
Histogram
Width Jitter
measure adjacent cycle deviation
Dp@lv
Statistics On
or use
Histogram
Cycle-to-Cycle
Jitter
count number of edges in a
waveform
edge@lv
--- ---
measure duty cycle
duty@lv
Statistics On
or use
Histogram
Duty Cycle Jitter
measure time interval error
tie@lv
Statistics On
or use
Histogram
Interval Error
Jitter
measure n-cycle
n-cycle@lv
---
N-Cycle Jitter
measure skew
skew
---
Clock
Skew
measure setup
setup
---
Setup
measure hold
hold
---
Hold
How to Use the Trend Tool
The Basic Idea
The Trend statistical tool displays the evolution of a timing parameter over time, in the form of a line graph. The
graph’s vertical axis is the value of the parameter; its horizontal axis is the order in which values are acquired.
x
Display the waveform to be analyzed.
x
Apply a timing parameter: period at level (p@lv), for example.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...