WavePro 7Zi
479
WP700Zi-OM-E-RevA
Example
ltpt
Local Time Peak-to-Trough
Definition
The time between a local feature peak and trough.
Description
The average of the time between all local feature peaks and troughs is displayed as ltpt.
Provided for histograms are the times between peak-trough pairs for all local features
between the parameter cursors.
Parameter
Settings
Selecting this parameter from the
Measure
menus prompts you for a hysteresis setting, which
allows you to set the hysteresis value to a specified number of vertical divisions.
Example
ltot
Local Time Over Threshold
Definition
The time a local feature spends over a user specified percentage of its peak-to-trough
amplitude.
Description
The peak-to-trough height of a local feature is measured. The time the local feature spends
over a user specified percent of the peak-to-trough height is then determined. The average for
all local features in a waveform is displayed as ltot. Provided for histograms is the time spent
over the threshold by each local feature between the parameter cursors.
Parameter
Settings
Selecting this parameter from the
Measure
menus displays hysteresis and level percent
menus. The associated fields allow you to set values in those menus: a specified number of
vertical divisions or a percentage of the peak-to-peak height of the local feature.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...