![LeCroy DDA 7 Zi series Скачать руководство пользователя страница 136](http://html1.mh-extra.com/html/lecroy/dda-7-zi-series/dda-7-zi-series_operators-manual_1866310136.webp)
WavePro 7Zi
135
WP700Zi-OM-E-RevA
Icon:
Description:
in the
LabNotebook
dialog. You can now create a hardcopy report of it, and email or print
it out.
Recalling Notebook Entries
After a notebook entry is made, you can recall it at any time. The recall includes waveforms and scope settings.
1.
Select the notebook entry from the list box.
2. Touch
Flashback
.
3.
To exit Flashback, touch the
Undo Flashback
button
in the top-right corner of the
screen, or press the
Auto
trigger button.
Note:
The flashback feature currently recalls the DSO Setup, and all displayed waveforms. Some forms of
‘result data’ are not recalled, including:
a. Persistence data. This will be saved in the hardcopy, and will be printed in the report, but will not be recalled
during Flashback.
b. Histogram data. Histograms internally have a 32-bit resolution, but when stored into a trace file and recalled
during flashback they are clipped to 16-bits.
c. Floating point waveforms. Certain math operations result in the creation of floating point waveforms with much
higher resolution than can be stored in a 16-bit waveform file. This extra resolution will not be preserved when
traces are recalled using flashback.
d. Cumulative Measurements. Any measurements that are on when the Lab Notebook entry is created are not
saved individually in the database, other than being embedded in the hardcopy image. This means that when
flashback is used, the measurements will be recomputed using the waveform data that was recalled. Normally
this will not pose a problem, but if cumulative measurements were on, which accumulated data from multiple
acquired waveforms, they will loose their history and show instead only the results from the stored waveforms.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...