![LeCroy DDA 7 Zi series Скачать руководство пользователя страница 290](http://html1.mh-extra.com/html/lecroy/dda-7-zi-series/dda-7-zi-series_operators-manual_1866310290.webp)
WavePro 7Zi
289
WP700Zi-OM-E-RevA
SETTING nT
For BES, EES, BEES, BESS, and EESS, this specifies the pit of interest. The results will be
computed for each space/pit (pit/space) pair using the subject pit and all the spaces within the
range specified.
Specifies the range of n indices that define the pits/spaces used in the calculation. The range
of n coupled with T are used to categorize the pits/spaces based on their widths.
Maximizing Performance
A basic guideline that you should follow to maximize the performance of calculation in multiple parameter
configurations is that precisely the same value for the clock period T, Threshold level, and Hysteresis value
should be used.
Following this guideline ensures that parameters can make use of results obtained in previous parameter
calculations. In most cases there is no need for different configurations of the above three items in different
parameter setups.
Pit or Space Identification
This is determined uniquely by the threshold, hysteresis, and edge polarity of threshold crossings. A positive
threshold crossing indicates the start of a positive polarity pit and the end of a negative polarity space. A positive
threshold crossing followed by a negative threshold crossing fully delineates a pit. A negative crossing followed by
a positive crossing fully delineates a space, as illustrated in the following figure.
In order to prevent false pit and space identifications, hysteresis is provided. Hysteresis adds an additional
condition that must be met before a threshold crossing is recognized as a pit/space edge. It requires that the
waveform make an excursion of a certain distance from the threshold before the next threshold crossing is
recognized.
The next figure shows a threshold crossing that would result in incorrect pit identification without hysteresis.
The hysteresis band shown in the next figure is centered on the user- selected voltage level threshold.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...