Determining Time Parameters
Time parameter measurements such as width, period and delay are carried out with respect to the mesial
reference level (see figure), located halfway (50%) between the top and base reference lines.
Time-parameter estimation depends on the number of cycles included within the observation window. If the
number of cycles is not an integer, parameter measurements such as rms or mean will be biased. However, only
the last value is actually displayed, the mean being available when statistics are enabled. To avoid these bias
effects, cyclic parameters can be chosen, including crms and cmean, that restrict the calculation to an integer
number of cycles.
Determining Differential Time Measurements
The instrument enables accurate differential time measurements between two traces: for example, propagation,
setup and hold delays (see figure).
If included in your oscilloscope, parameters such as Delta c2d± require the transition polarity of the clock and data
signals to be specified.