Restoring Software
System Recovery for Oscilloscopes Running Windows XP and Vista
Your oscilloscope is designed to operate reliably for many years. However, the application software operating the
instrument runs on a Windows platform. The loading or incomplete removal of additional Windows applications
may eventually cause problems in the stability of the operating system. Severe cases may require a reloading of
the base operating system and oscilloscope application. This is done using a recovery routine that restores a
clean copy of the image originally installed on the C: drive.
Note
: Any user and calibration data located on the D: partition is not affected by the recovery process.
LeCroy provides you with a recovery application and a backup image in an extra partition on the instrument’s hard
drive. The recovery process is easy to perform, using the following instructions . After the recovery procedure is
done, you must activate Windows, either by Internet connection to Microsoft’s Web site or by telephone. Have
your Windows Product Key number (located on the rear of the oscilloscope) handy during Widows reactivation.
Note
: The recovery process produces a replica of the operating system and oscilloscope application software at
the current revision levels when the oscilloscope was manufactured. Any further revisions of the application
software, Windows operating system, and virus scan definition files are not automatically upgraded. Therefore,
after completion of the disk image recovery, it is highly recommended to search vendor Web sites and upgrade
the individual components to current revision levels.
The current oscilloscope application software can be downloaded directly from the LeCroy website at
www.lecroy.com
. Since the calibration data for the oscilloscope is stored in the D: drive, current calibration
constants are not overwritten during the recovery process.
Recovery Procedure
1.
Connect a network cable to the LAN port on the rear of the oscilloscope if you intend to activate windows
through the Internet.
2.
Connect a keyboard and a mouse to the oscilloscope.
3.
Apply power to the oscilloscope.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...