WavePro 7Zi
195
WP700Zi-OM-E-RevA
Note
: Depending on whether you select a vertical or horizontal parameter, the correct units will be automatically
displayed (V, s, Hz, dB) in the
Between
and
And
fields. Or, if you select a simple ratio parameter (such as power
factor) that yields a dimensionless number, no units will be displayed.
6. Touch
the
Find Range
button to quickly display the most recent value of the parameter measurement.
Waveform Gated Parameters
Setting Up Waveform Qualifiers
1. Touch
Measure Measure Setup
.
2. Touch
a
Px
tab to open the setup dialog.
3. Touch
inside
the
Source
field and select a source from the pop-up menu.
4. Touch
inside
the
Measure
field and select a parameter from the pop-up menu.
5. Touch
the
Accept
tab of the dialog on the right, then touch the
Values Based on Waveform State
checkbox.
6. Touch
inside
the
When Wform
field and select the gating source.
7. Touch
inside
the
State Is
field and select
High
or
Low
from the pop-up menu. Parameter measurements
on the subject waveform will only be taken when the gating waveform is in the selected state.
8. Touch
inside
the
Level Type
field and select
Absolute
or
Percent
from the pop-up menu.
9. Touch
inside
the
Level
field and enter the crossing level value at which you want measurements to begin
using the scroll bar at the bottom of the window. Touch the keypad button
to enter a value using
the pop-up keypad.
OR
Touch the
Find Level
button to automatically select the 50% level of your gating waveform.
EMC Parameters
Not available for all scopes, the EMC Pulse Parameter software package includes enhanced Rise@level,
Fall@level, Width@level, Time@level, and DTime@level parameters. The functionality includes user definable
thresholds for accurate pulse measurements.
% pk-to-pk
is for ESD pulses that do not start at 0; they might go from negative to
positive or positive to negative voltage. The risetime is measured as a percentage of
peak to peak not base to top.
% 0-to-max
is for ESD pulses that start at 0 V and reach a peak positive voltage. The
risetime is measured as a percentage of 0 to max instead of base to top of a traditional
IEEE pulse.
% 0-to-min
is for ESD pulses that start at 0 V and reach a peak negative voltage. The
risetime is measured as a percentage of 0 to min instead of top to base.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...