WavePro 7Zi
51
WP700Zi-OM-E-RevA
Pass/Fail Testing
Comparing Parameters
Each Pass/Fail input (
Qx
) can compare a different parameter result to a user-defined limit (or statistical range)
under a different condition.
The conditions are represented by these comparison operators:
At the touch of a button, test results can also be compared to these standard statistical limits:
In Dual Parameter Compare mode, your X-Stream oscilloscope can take parameter results measured on two
different waveforms and compare them to each other. If desired, set your test to be true if
Any
or
All
waveforms
fit the criteria set by the comparison condition. Setups are conveniently shown in the Summary box of the
Qx
dialog as follows:
Mask Tests
Mask testing can be done using an existing mask, or by using a mask created from your actual waveform, with
vertical and horizontal tolerances that you define. Existing masks can be loaded from a floppy disk or from a
network.
You can set your mask test to be True for waveforms All In, All Out, Any In, or Any Out. For example, if you select
All In, the test will be False if even a single waveform falls outside the mask.
Masks that you create from your waveform can be confined to just a portion of the trace by use of a measure
gate. Refer to the
Measure Gate
topic for more information on how this feature works.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...