![LeCroy DDA 7 Zi series Скачать руководство пользователя страница 231](http://html1.mh-extra.com/html/lecroy/dda-7-zi-series/dda-7-zi-series_operators-manual_1866310231.webp)
Operator's Manual
WP700Zi-OM-E-RevA
230
2.
In the Sampling Mode area, select
Sequence
mode. A tab labeled “Sequence” will appear
next to the Timebase tab. Touch that tab.
3.
In the Sequence tab, select a
display mode
.
Select the number of segments to be displayed
at one time.
4.
If you have acquired more segments than you
can display at one time, you can choose the
segment to begin the display at.
5.
Set up the Serial Trigger to capture the event you desire. For instance, you might want to trigger on a
specific address or data value, and capture long pre-trigger time to determine what precedes that
message. In this example, we’ve used a simple I
2
C Start trigger. To begin the sequence mode acquisition,
press the front panel
S
INGLE
trigger button. Each time the trigger condition is met, the TRIG’D light on the
front panel will flash. When you’ve acquired the set number of segments, the trigger will STOP and the
display will appear as below (this is a 50 segment acquisition in Mosaic display mode).
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...