![LeCroy DDA 7 Zi series Скачать руководство пользователя страница 389](http://html1.mh-extra.com/html/lecroy/dda-7-zi-series/dda-7-zi-series_operators-manual_1866310389.webp)
Operator's Manual
WP700Zi-OM-E-RevA
388
7. Touch
the
Trend
button
at the bottom of the dialog. A
Math Selection for Trend
pop-up
menu
opens from
which you must select a math trace to display the trend.
8.
A second setup dialog opens to the right of the main with more setup options. The options offered depend
on the parameter you chose, but all include
Level is
,
Percent Level
,
Slope
, and
Hysteresis
. A
Find
Level
button is also provided in this mini-dialog.
Option Field
Settings
Level Is
Percent
Absolute
Percent Level
0 to 100%
Slope
Positive
Negative
Hysteresis
0 div to 10 div
Note:
The
Hysteresis
selection imposes a limit above and below the Level, which precludes measurements of
noise or other perturbations within this band. The width of the band is specified in milli-divisions.
Guidelines for Use
1. Hysteresis must be larger than the maximum noise spike you want to ignore.
2. The largest value of hysteresis usable is less than the distance from the level to the closest extreme value of
the waveform.
3. Unless you know the largest noise and closest extreme level that will ever occur on any cycle, leave some
margin on both sides of the level.
Math Setup
Now that the parameter setups are done, you have to set up the Trend math function.
1. Touch
Math
in the menu bar, then
Math Setup...
in the drop-down menu.
2.
In the "Math" dialog, touch the
Fx
tab
for the math trace you chose to display the trend.
3.
Touch the "Trend" tab
in the setup dialog to the right of the main setup dialog. Touch
inside the
Values to Trend
data entry field and enter a value from 20 to 20,000, using the pop-up numeric
keypad.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...