Operator's Manual
WP700Zi-OM-E-RevA
164
menu
.
5.
In the mini-dialog to the right of the main setup dialog, touch the
Gate
tab to narrow the span of
measurement.
6. Touch
the
Accept
tab to define parameter values to be used in the measurement.
Parameter Status
Status Symbols
Below each parameter appears a symbol that indicates the status of the parameter, as follows:
Symbol: Description:
A warning symbol indicates that there is something wrong with the signal or the setup. Touch the
parameter list box and read the explanation in the message line at the bottom of the screen.
A green check mark means that the scope is returning a valid value.
A crossed-out pulse means that the scope is unable to determine top and base; however, the
measurement could still be valid.
A downward pointing arrow indicates an underflow condition.
An upward pointing arrow indicates an overflow condition.
An upward-and-downward pointing arrow indicates an underflow and overflow condition.
Using X-Stream Browser to Obtain Status Information
Example:
Here is a case of an overflow condition, in which the amplitude of the waveform cannot be determined:
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...