Operator's Manual
WP700Zi-OM-E-RevA
458
Sector Pulse
Sector Pulse triggers on the specified (
n
th
) sector start pulse after the Index mark. To set it up, proceed as follows:
1.
Press the
Disk Drive Analysis (DDA)
button on the Quick Access toolbar. The disk drive dialogs appear.
2. Touch
the
Disk Triggers
tab.
3.
In the "Disk Trigger Types" menu
, touch
Sector Pulse
. The Sector Pulse data
entry fields appear. See
Sector Pulse
in "Setting Up to Use Drive Channel Analysis then touch
Back
to complete this setup.
4. Touch
inside
the
Trigger On Sector x after Index
data entry field and enter a value from 0 to 999 using
the pop-up numeric keypad.
5.
Touch inside the Sector Pulse
Source
field and select an input channel or external source from the pop-up
menu.
6.
Touch inside the Sector Pulse
Polarity
field and select positive or negative polarity from the pop-up menu.
7.
Touch inside the Sector Pulse
Level
field and enter a value from -250 mV to 250 mV using the pop-up
numeric keypad.
8.
Touch inside the Index
Source
field and select an input channel or external source from the pop-up menu.
9.
Touch inside the Index
Level
field and enter a value from -250 mV to 250 mV using the pop-up numeric
keypad.
Servo Gate
Servo Gate triggers on all or selected Servo Gate pulses starting after the Index mark. To set it up, proceed as
follows:
1.
Press the
Disk Drive Analysis (DDA)
button on the Quick Access toolbar. The disk drive dialogs appear.
2. Touch
the
Disk Triggers
tab.
3.
In the "Disk Trigger Types" menu
, touch
Servo Gate
. The Servo Gate data entry
fields appear.
4. Touch
inside
the
After index wait x servo gates
field and enter a value from 0 to 999 using the pop-up
numeric keypad.
5. Touch
inside
the
then skip x gates between triggers
field and enter a value from 0 to 999 using the pop-
up numeric keypad.
6.
Touch inside the Servo Gate
Source
field and select an input channel or external source from the pop-up
menu.
7.
Touch inside the Servo Gate
Polarity
field and select positive or negative polarity from the pop-up menu.
8.
Touch inside the Servo Gate
Level
field and enter a value from -250 mV to 250 mV using the pop-up
numeric keypad.
9.
Touch inside the Index
Source
field and select an input channel or external source from the pop-up menu.
10. Touch inside the Index
Level
field and enter a value from -250 mV to 250 mV using the pop-up numeric
keypad.
Setting Up Zoom
You can zoom on a particular segment and byte of your signal as follows.
1. Press
the
Disk Drive Analysis (DDA)
button on the Quick Access toolbar. The disk drive dialogs appear.
2. Touch
the
Zoom
tab.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...