![LeCroy DDA 7 Zi series Скачать руководство пользователя страница 297](http://html1.mh-extra.com/html/lecroy/dda-7-zi-series/dda-7-zi-series_operators-manual_1866310297.webp)
Operator's Manual
WP700Zi-OM-E-RevA
296
Ending Edge Shift Sigma cannot be calculated for a given index
n
unless there are at least two Ending Edge Shift
values calculated for that n index.
When Ending Edge Shift is configured as a custom parameter with a range of
n
, the value calculated is the
standard deviation of the distribution that results by normalizing each independent distribution categorized by the
space (pit) nT following the subject pit (space). Distributions are normalized by subtracting the mean of the
distribution from all of the elements in the distribution. This results in the following equation for overall Ending
Edge Shift Sigma resulting from the individually categorized Ending Edge Shift Sigma values:
Note:
The value calculated by EESS will generally not be the same as the sigma of EES measurement when a
range of
n
is used and statistics are on. This is because the two measurements are not the same. The EESS
measurement normalizes the results for each
n
by subtracting the mean EES from each EES in the nth
distribution. This results in a superposition of mean-centered distributions, not a superposition of 0-centered
distributions contributing to EES measurements. EESS will always be less than or equal to the standard
deviation of EES measurements.
Display Options
ORM parameter calculations can be displayed, histogrammed and trended in a variety of ways.
DISPLAY TYPE
VALUE DISPLAYED
Parameter
Statistics Off
Single value of the standard deviation of the mean normalized ending edge shift values for
pits/spaces of interest for last acquisition.
Parameter
Statistics On
average, minimum, maximum, and sigma of the ending edge shift sigma value calculated
per acquisition for all acquisitions since the last CLEAR SWEEPS operation.
nT Table
List of values of the standard deviation of the ending edge shift values for each ‘nT’ spaces
(pit) within the selected range following the subject pit (space) for the last acquisition.
Histogram
Function
Histogram of ending edge shift sigma values calculated for each acquisition for all
acquisitions since the last CLEAR SWEEPS operation.
Trend Function
Trend of the ending edge shift sigma values calculated for each acquisition for all
acquisitions since the last CLEAR SWEEPS operation.
Beginning Ending Edge Shift (BEES)
BEES provides a measurement of both the beginning and ending edge shift for a subject
n
pit (space) preceded
and followed by a specified space (pit). (See
BES
and
EES
.) The measurement is calculated between the points
where the data and clock signals cross selected voltage thresholds. The clock period T can be entered by the
user, or measured from a user supplied clock signal, as described below.
The value calculated depends on the clock and data edges selected, as shown in the table below. The
Data
Slope
menu selects the polarity of the
subject n
pit/space. If
Pos
(positive) is selected, the measurement is
performed from the beginning and ending edges of positive polarity pits and is preceded and followed by a space
of the specified width. If
Neg
(negative) is selected, the measurement is performed from the edges of negative
polarity spaces and is preceded and followed by a pit of the specified width.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...