WavePro 7Zi
371
WP700Zi-OM-E-RevA
Polarity
In many electrical standards, such as DS-1, alternate “ones” are inverted. Each time a one is transmitted it is
either a positive or negative going pulse depending upon the polarity of the previous one. This type of coding is
referred to as AMI (alternate mark inversion). The
Polarity
control allows you to select which polarity (positive or
negative) pulse to test.
The STS-3E and STM-1E standards use CMI (code mark inversion) pulse coding. In CMI coding, a one remains
high for the full bit period while a zero has a transition to the low state in the middle of the bit period. The Polarity
control allows you to select whether a 1 or 0 is to be tested.
Attenuation
The
Attenuation
control allows you to enter an attenuation value to be applied to the test signal. This control
allows for amplitude correction for cable or other systematic losses. The attenuation value can be set from 0.5 to
1 in steps of 0.01, with 1 meaning “no external attenuation” and 0.5 corresponding to an amplitude reduction of
50%. If a cable with 3 dB of loss were being used to couple the signal under test to the instrument, for example,
the attenuation value should be set to 0.708.
Offset
The offset control allows for the correction of DC offsets in the signal under test. This control can be varied from -
50 mV to 50 mV.
Clear Sweeps and Pause
The Clear Sweeps button allows you to reset the sweep count and start testing over again. The Pause button,
which suspends testing without resetting the sweep count, becomes a Run button once it is pressed. Pressing the
Run button, in this case will start the test from where it was last suspended.
Actions
The Actions section contains a set of check boxes that determine what actions are taken when a test fails. Actions
are selected by checking the appropriate box; multiple actions are possible for each failure. The available actions
are:
Save
- Stores the failed waveform in a file. When this option is selected, the Save Waveform tab appears in the
main dialog. Touch this tab to set the file name and storage location.
Stop
- Stops testing on the first failure.
Alarm
- Generates an audible alarm on each failure.
Pulse
- Generates a pulse at the auxiliary output BNC connector on the front panel of the instrument. When this
option is selected, the Aux Output tab appears. Touch this tab to set the type of pulse.
Hardcopy
- Prints a hard copy image of the failed mask test. Hard copies can be sent to any valid printer (either
networked or directly connected to the instrument). It is also possible to print to file or email the image. When this
option is selected, the Hardcopy tab appears. Touch this tab to set the format and devices.
Stop Test
- Gives you the ability to stop the test after a predetermined number of sweeps have been completed.
When this option is selected, the After field alongside the checkbox becomes active. Touch inside the After field to
set the number of sweeps, by means of the pop-up keypad.
The
Force Actions
Once button immediately executes all of the selected actions when it is pressed.
Adding Measurements to a Mask Test
In addition to measuring pulse compliance relative to a given mask, parametric measurements are also available.
A full set of measurements including amplitude, overshoot and rise time are available from the oscilloscope’s
Measure menu. Measurements are made on the displayed waveform as it appears on the mask test screen.
Setting up Parametric Measurements
1.
In the menu bar, touch Measure; then Measure Setup in the drop-down menu. The menu shown below is
displayed:
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...