WavePro 7Zi
283
WP700Zi-OM-E-RevA
To Configure a Trend:
1.
From the menu bar select
Math
, then
Math Setup…
from the drop-down menu.
2. Touch
an
Fx
tab that is not currently assigned a math function (i.e., Zoom function by default).
3. Touch
inside
the
Source1
field and select a source waveform from the pop-up menu.
4. Touch
inside
the
Operator1
field and select Trend from the
Select Math Operator
pop-up menu. The
"Trend" setup dialog will appear at the right of the screen:
5.
Decide whether all the parameter values, all per trace, or only the average of all parameter calculations for
each waveform acquisition should be placed in the trend.
All
- every parameter calculation on each waveform will be placed in the trend.
Average
- trends only the average of all the values calculated on a given acquisition and yields one point
in the trend per acquisition.
All per Trace
- for each acquisition, clears the buffer and places all parameter calculations from the new
data in the trend. Unless this is specifically required,
All
should be selected.
6.
Choose the number of values to be placed in the generated trend.
7.
If desired, you can also configure the center and height of the trend in the base units of the parameter
being trended. However, this is not a requirement, and
Find Scale
can be used to center the trend after it
has been calculated.
Center
is for selecting the mantissa, exponent, or number of digits resolution. The configuration is the
value at the horizontal center line on the grid, while units are those of the parameter trended.
Height/div
selects the vertical value of each vertical screen division. Units are those of the parameter
trended.
Trend Calculation
Once the trend has been configured, parameter values will be calculated and trended on each subsequent
acquisition. Immediately following an acquisition, its trend values will be calculated. The resulting trend is a
waveform of data points that can be used the same way as any other waveform. Parameters can be calculated on
it, and it can be zoomed, serve as the
x
or
y
trace in an XY plot, and can be used in cursor measurements.
The sequence for acquiring trend data is:
1. trigger
2. waveform acquisition
3. parameter calculations
4. trend update
5. trigger rearm
If the timebase is set in non-segmented mode, a single acquisition occurs prior to parameter calculations.
However, in segment mode, an acquisition for each segment occurs prior to parameter calculations. If the source
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...