WavePro 7Zi
443
WP700Zi-OM-E-RevA
Eye Diagram Setup and Measurements
1.
Select Mask for each Eye Diagram -- Depending on the Standard selected, there may be one or several
types of serial data masks available.
2.
Select a Mask Margin (as % of the nominal mask size) if desired.
3.
Select a Measurement Type. There are three groups of parameters:
x
Amplitude
Eye
Height
One Level (Eye)
Zero Level (Eye)
Eye
Amplitude
x
Timing
Eye
RiseTime
Eye
FallTime
x
Eye
Eye
Width
Eye RMS Jit
Eye
BER
Eye
Crossing
Avg Pwr (Eye)
Mask Test (Mask Violation Locator) Setup
1.
Select the "Mask Violation Locator" tab.
2. Check
the
Testing On
checkbox to detect mask violations.
3. Check
the
Show Location
checkbox to select display of bits where mask violations occur.
4. Check
the
Stop on Error
checkbox after
N Failures
.
Figure 5-48. Mask Violation Setup.