Operator's Manual
WP700Zi-OM-E-RevA
282
increase beyond that which can be
displayed on screen using the current
vertical scale, this scale is automatically
increased in a 1-2-5 sequence.
x
LinConstMax
sets the vertical scaling to
a linear value that uses close to the full
vertical display capability of the scope.
The height of the histogram will remain
almost constant.
Setting Binning and Histogram Scale
For either the
Linear
or
LinConstantMax
vertical scale option, the scope automatically increases the vertical
scale setting as required, ensuring that the highest histogram bin does not exceed the vertical screen display limit.
The
Center
and
Width
fields allow specification of the histogram center value and width per division. The width
per division multiplied by the number of horizontal display divisions (10) determines the range of parameter values
centered on the number in the
Center
field, used to create the histogram.
Displaying Trends
The Trend function for processing waveforms creates a graph of successive waveform parameter values. It
provides useful visual information on waveform parameter variation. Used together with other scope features, it
allows you to graph certain parameters compared to others.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...