![LeCroy DDA 7 Zi series Скачать руководство пользователя страница 457](http://html1.mh-extra.com/html/lecroy/dda-7-zi-series/dda-7-zi-series_operators-manual_1866310457.webp)
Operator's Manual
WP700Zi-OM-E-RevA
456
11. Touch inside the
Boost
field and enter a value from 0 to 13 dB using the pop-up numeric keypad. See
Boost at fc
in "PRML Channel Emulation" then touch
Back
to complete this setup.
12. Touch inside the
Group Delay
field and enter a value from –30.0% to +30.0% using the pop-up numeric
keypad. See
Group Delay
in "PRML Channel Emulation." This parameter is not set by the "Train Filter"
function.
Setting Up Analog Compare
Drive Analysis Setup
1. Press
the
Disk Drive Analysis (DDA)
button on the Quick Access toolbar. The disk drive dialogs appear.
2. Touch
the
Drive Analysis
tab
.
3. Touch
the
Analog Compare
button
. The Analog Compare setup checkbox, buttons, and
data entry fields appear.
4. Touch
inside
the
Head Signal
field and select a signal source from the pop-up menu. The choices
comprise channel inputs C1-C4, math traces F1-F8, memories M1-M4, or Reference. See
Using Analog
Compare
for guidelines on using a Reference.
5. Touch
the
Trace On
checkbox to turn the Head Signal on.
6. Touch
inside
the
Read Gate
field and select a source from the pop-up menu. The choices include
Ref
and
none
. Read GateIf the Read Gate signal is connected to a DDA-5005 channel and specified, it will be used
to determine the regions of the signal to be analyzed. If Read Gate is not present, the entire waveform will
be used unless the Analyze Region cursors are enabled.
7.
If for Read Gate you selected other than
none
, Touch inside the
Gate Polarity
field and select positive or
negative polarity.
8. Touch
the
Store Head Reference
button.
9. Touch
the
Specify Region
checkbox to specify a start and end time, if desired. (This may be necessary if
you are not using Read Gate. If you are not using Read Gate, the Analysis Region must start with a
preamble for VCO synchronization.) Then touch inside the
Start
and
End
data entry fields and enter
starting and ending time values from -1.0 ks to +1.0 ks, using the pop-up numeric keypad.
10. To jump to a position in the head signal, touch inside the "Position"
Segment
field and enter a value from 1
to 999, using the pop-up numeric keypad. Then touch inside the
Byte
field and enter a value from 50 to
50,000.
11. Touch inside the
Worst Error #
field and enter a value from 1 to 100, using the pop-up numeric keypad.
12. Touch inside the
Analog Threshold
field and enter a value from 0 mV to 1 V.
Channel Setup
Channel Setup is required for
Analog Compare
, which is the same as for
Channel Emulation
.
Setting Up Noise Analysis
1.
The disk drive dialogs appear.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...