140
Setting Up a Math Function
Math Setup
This setup mode allows you to quickly apply frequently used math functions.
1. Touch
Math
Math Setup
on the menu bar.
2.
If there are math functions already assigned to
F1
through
Fx
[The number of math traces available
depends on the software options loaded on your scope. See Specifications.] , touch the checkbox for the
function you want to enable.
3.
To assign a new math function to a trace, touch the Fx button for that trace, for example
. The math
function menu appears.
4.
Touch a menu selection; your new function is automatically assigned, with the same setups as were in
place for the last function in that Fx position.
5.
If you want to change other setup items, like the source waveform, touch the appropriate Fx tab, for
example
. The setup dialog for that Fx position appears.
6. Touch
the
Single
function button
if you want to perform just one math function on the trace, or
touch the
Dual
function button
to perform math on math.
7. Touch
the
Graph
button, then touch inside the
Graph with
field to select a graph mode. The Graph modes
are as follows:
Mode:
Description:
Histogram of the values of a parameter
Track of each value of a parameter
Trend of the values of a parameter over time
Both Track and Trend are tools that can be used to plot measurement data and observe variations with respect to
time. Differences between Track and Trend are summarized in the table below:
Characteristic
Trend
Track
Representation
Parameter value vs. event
Parameter value vs. time
Behavior
Cumulative over several acquisitions up to 1
million events
Non-cumulative (resets after
every acquisition). Unlimited
number of events
Time Correlation to Other
Data
No Yes
Monitors an Evolution in
the Frequency Domain
No. Trend points are not evenly spaced in time
and therefore cannot be used for an FFT.
Yes
Monitors the Evolution of a
Measurement Parameter
over Several Acquisitions
Yes
No. Track resets after every
acquisition.
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...