Operator's Manual
WP700Zi-OM-E-RevA
326
In math setup,
F1
is defined as the trend of the
ptop
parameter and
F2
as the trend of the
pwid
parameter.
Because we are expecting 1800 pits/spaces, make sure that for each math setup the trends are set to use up to
20,000 values.
Note:
If configuration tracking is used on the ptop parameter, the pwid parameter must be visited in order to set
the polarity to positive because ptop inherently implies positive polarity pits.
The trigger is set to trigger on a pit edge and is operated initially in single-shot mode. For convenience, the
waveforms are ordered on the screen in a particular manner so that they will automatically work correctly with XY
display mode:
1. F2: Trend of
t@pit
2. F1: Trend of
pwid
3. Channel 1: optical recognition data signal
Note:
Prior to acquisition, select each trace and press the RESET button to ensure that all the traces are reset.
The single-shot trigger button is pressed and the waveform acquired. The waveform should be centered on the
screen. The trends will typically have data not well centered or off-screen. These traces can be positioned on the
grid by touching the
F
IND
S
CALE
button in each trend setup dialog. The screen shown here is what each trace looks
like after the waveform has been acquired and the trends centered.
The waveforms are displayed in Quad grid mode. Notice that
there are exactly as many events inside both trends, a
necessary condition. Although the trends are very short
(containing only 902 out of the total 20,000 pits allowed)
repeated triggering will eventually fill in both trends sufficiently.
Set the display mode to
XY
. Clusters of pit top values will be apparent: clustered because the tops tend to be
approximately the same amplitude and the pit widths approximate multiples of the clock period. Select
F2
(the
trend of pwid) and use the vertical
Z
OOM
control to expand the X-axis scale. Select
F1
(the trend of ptop) and use
the
Z
OOM
knob to adjust the vertical scale. The vertical
P
OSITION
knob can be used to position the display vertically.
This is what the XY plot looks like:
Содержание DDA 7 Zi series
Страница 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Страница 2: ... L R R H HUD RU D D ...
Страница 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Страница 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Страница 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Страница 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Страница 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...