Assembly and preparation
92
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Assembly overview
This chapter describes the following basic
assembly tasks:
– Docking/undocking the M540 from the M500
– Locking/unlocking the M540 into the M500
– Connecting/disconnecting the system cables
IACS components are compatible with
commercially available mounting solutions.
Commercially available mounting
solutions
Various mounting solutions are available. It is the
responsibility of the hospital to install, test, and
ensure the proper and safe operation of any
mounting solution.
Contact your Dräger representative for specific
approved mounting solutions.
CAUTION
Avoid mounting solutions that could impede air
flow since the M500, PS250 / P2500 require ade-
quate airflow to dissipate heat. In addition, when
mounting the PS250 / P2500, always position it
vertically for adequate heat dissipation.
CAUTION
Check the weight ratings of the commercially
available mounts to avoid injuring the patient or
damaging the device.