![Dräger Infinity Acute Care System M540 Скачать руководство пользователя страница 552](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773552.webp)
Troubleshooting
552
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
None
%0 out of range
high
1)
%0 out of range
low
1)
Parameter value
The C.O. is greater
than 20 liters/min
or less than
0.5 liters/min
because of:
– a physiological
condition
– unstable
baseline
– incorrect
injectate
volume,
catheter size,
or computation
constant
– faulty catheter,
cable, or
cartridge
– Check the patient
and treat if
necessary.
– Use cooler
injectate.
– Enter the correct
values in the C.O.
page.
– Repeat the
measurement. If
message persists,
replace faulty
components.
None
%0 Check Injectate
Probe
Parameter value
The thermistor is
not connected or
became
disconnected
during a
measurement.
Connect the probe
and repeat the
measurement.
None
%0 duplicate device
connected
1)
Parameter value
Multiple C.O.
sources are
connected. This
includes CCO
devices connected
via the device
connectivity
option.
Disconnect duplicate
C.O. sources.
None
%0 Injectate Too
Cold
1)
Parameter value
The injectate
temperature is too
cold during the
measurement
process.
– Use an injectate
within the correct
temperature
range of –5 °C to
+30 °C (–23 °F to
+86 °F).
– Check equipment
and replace if
necessary.
1)
%0 is a placeholder for the parameter label C.O.
Priority
Message
Parameter field
Problem
Solution