Troubleshooting
526
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
SpO2
The following messages originate from three
different hardware devices (Masimo SET, Masimo
rainbow SET, and Nellcor OxiMax).
Priority
Message
Parameter field
Problem
Solution
None
(Masimo
rainbow
SET only)
Learning pulse CO-
Ox
Learning
Displays
parameter values
for SpO2, PLS,
and PI. Parameter
values for SpHb
(SpHbv), SpOC,
SpMet, PVI, SpCO
are replaced by ***
The parameters have
been detected but
have not yet been
computed.
Wait until message
disappears.
None
(Masimo
rainbow
SET only)
Low %0 SIQ
1)
Associated
parameter values
are still displayed
– Poor signal quality
– Measurement
reading is
obscured
– Check the
patient and
treat if
necessary.
– Make sure the
SpO2 sensor is
attached
properly to the
patient.
– Check all cable
connections.
None
Low SpO2 SIQ
Low SpO2 SIQ
Parameter values
are still displayed.
The Masimo MCable
detects low signal
quality
– Check the
patient and
treat if
necessary.
– Make sure the
SpO2 sensor is
attached
properly to the
patient.
– Check all cable
connections.
1)
%0 is a placeholder for the following parameter labels: PVI, SpHb (SpHbv), SpMet, SpOC, SpCO.