External devices – Evita 2D, Evita 4, Evita XL (Medibus)
434
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Supported Medibus ventilator data
The Evita ventilators (Evita 2D, Evita 4, and Evita
XL) send parameters, settings, modes and alarm
messages to the Cockpit.
Certain ventilator parameters, settings, and
ventilation modes are broadcast to the Infinity
network. In addition, a limited number of alarms are
made available to the Infinity network.
Refer to the RS-232 export handbook for
information on which parameters are available for
export protocol and which settings and modes are
supported. The RS-232 export handbook is
available in English only.
Supported Medibus parameters on the Cockpit
The following table lists which parameters are
displayed and trended on the Cockpit. Refer to
page 433 for a list of supported ventilators and the
compatible software versions.
Parameter
label
Description
Unit of
measure-
ment
Displayed in
Ventilation
Show
all
page (see
page 438) Yes/No?
Cockpit
Trends
page
Available
for display
in
parameter
field
Yes/No?
Cdyn
Dynamic compliance L/bar
Yes
Continuous
No
etCO2
End-tidal carbon
dioxide
concentration
mmHg
Yes
Continuous and
mini-trend
Yes
kPa
%
FiO2
Inspiratory oxygen
fraction
%
Yes
Continuous
No
MVe
Minute volume,
expired
L/min
Yes
Continuous and
mini-trend
Yes
MVspon
Minute volume,
expired,
spontaneous
L/min
Yes
Continuous
No
P0.1
Occlusion pressure
mbar or
cmH2O
Not supported for
display.
Not trended
No
PEEP
Peak end expiratory
airway pressure
mbar or
cmH2O
Yes
Continous Yes
Pmean
Mean airway
pressure
mbar or
cmH2O
Yes
Continuous
Yes
Pmin
Minimum airway
pressure
mbar or
cmH2O
Yes
Not trended
No