Troubleshooting
Mainstream CO
2
Priority
Message
Cause
Remedy
None
CO
2
calibration check
failed
The mainstream sensor
calibration procedure failed.
Contact DrägerService
®
.
None
CO
2
calibration check in
progress
The mainstream sensor
calibration procedure is in
progress.
Informational message – no
action required.
None
CO
2
calibration check
successful
The mainstream sensor
calibration procedure was
successful.
Informational message – no
action required.
None
CO
2
calibration failed
The mainstream sensor
calibration procedure was
unsuccessful.
Try again or contact
DrägerService
®
.
None
CO
2
calibration in
progress
The mainstream sensor
calibration procedure is in
progress.
Informational message – no
action required.
None
CO
2
calibration
successful
The mainstream sensor
calibration procedure was
successful.
Informational message – no
action required.
!
%0 check airway
adapter
1)
– The mainstream sensor
is not properly seated on
the adapter
– There are secretions in
the adapter.
– There is sensor zero drift
– Make sure the mainstream
sensor is attached properly
to the adapter.
– If message persists, clean
or replace the airway
adapter.
– If message persists though
the airway adapter is clean,
zero the sensor.
!
%0 H/W failure
1)
CO
2
sensor hardware
failure.
Contact DrägerService
®
.