Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
259
SpO2 and Pulse CO-Ox monitoring with Masimo SET MCable
To apply the sensor
1
Select the size and type of sensor that is best
suited for your patient. Follow the
recommendations of the manufacturer.
2
Position the sensor correctly and attach it to
your patient.
3
Connect the sensor to the Masimo SET MCable
or the Masimo rainbow SET MCable.
SpO2
and
Pulse CO-Ox display
On the Cockpit, the SpO2 display consists of:
– SpO2 parameter field
– A user-configurable Pulse CO-Ox parameter
field when a Masimo rainbow SET MCable with
additional parameters activated is connected.
– SpO2 pulse plethysmogram waveform
The following table lists the maximum times the
M540 requires to report the parameter values after
connecting the sensor to the MCable.
33
0
NOTE
After connecting the sensor, if the sensor-LED
does not light up:
– observe the monitor for any message and act
accordingly, or
– replace the sensor.
NOTE
The pulse plethysmogram waveform is directly
proportional to the strength of the pulse
amplitude.
Parameter
Maximum time
SpO2, PLS, PI
Up to 35 s
SpMet, PVI, SpCO
Up to 60 s
SpHb, SpOC
Up to 90 s
PVI
Up to 150 s