![Dräger Infinity Acute Care System M540 Скачать руководство пользователя страница 346](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773346.webp)
Mainstream CO
2
monitoring
346
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Selection
Available settings
Description
Zero
(Only available if a
CO
2
device is
connected)
None
Zeroes the CO
2
sensor if necessary. The
CO
2
sensor stores a new isoelectric point for
CO
2
measurements.
Scale
– 0 to 40 mmHg (default), 0
to 60 mmHg,
0 to 80 mmHg, 0 to
100 mmHg
– 0 to 5 kPa (default), 0 to
8 kPa, 0 to 12 kPa, 0 to
16 kPa
– 0 to 5 % (default), 0 to 8 %,
0 to 12 %,
0 to 16 %
Adjusts the size of the CO
2
waveform.
Scale settings apply to both Mainstream and
Microstream.
Atm. pressure
570 to 800 mmHg
760 mmHg (default)
Determines the ambient pressure setting of
the sensor and compensates for pressure
effects. Failure to compensate for pressure
can cause inaccurate measurements.
Gas
compensation
Air
(default), N
2
O/O2,
O2>50%
,
HeliOx
Compensates for supplemental oxygen or
N2O or Heliox. Failure to compensate for
supplemental oxygen can cause inaccurate
measurements.
RRc apnea time
[s]
Off
(default), 10, 15, 20, 25,
30 s
Specifies the time the M540 waits before
reporting a cessation of breathing as an
apnea event.
RRc apnea time [s]
settings apply to both
Mainstream and Microstream.
RRc apnea
archive
Off
,
Store
(default),
Str/Rec
,
Record
Determines what happens in response to an
apnea.
RRc apnea archive
settings apply to both
Mainstream and Microstream.
Airway adapter
Reusable
(default),
Disposable
Determines the type of airway adapter used
for
CO
2
monitoring.
Compensates for the type of airway adapter
that is being used.
Requires matching the adapter with the
configuration setting at the Cockpit; if the
adapters do not match, the
CO
2
value
displayed is compromised.