ECG, arrhythmia, and ST segment
202
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Connecting the lead sets for 12-lead monitoring
The ECG lead sets connect directly to the M540.
A
M540 ECG port
B
6-wire lead set
C
4-wire lead set
To connect the ECG lead sets
1
Insert the 6-wire lead set (B) and the 4-wire lead
set (C) into the ECG port (A) on the side of the
M540.
Orient lead sets (B and C) so the exposed pins
face towards you as you push them firmly into
the channel.
2
Connect the lead wires to the patient. For
information on applying the electrodes to the
patient, refer to the figures starting on
page 208.
31
1
B
C
A
5
NOTE
An ECG lead set can rest in the ECG port of the
M540 without actually being connected. Make
sure that all ECG lead sets are pushed firmly into
the ECG port of the M540.
Almost every MonoLead features a number on
the lead set indicating how many leads connect.
When connecting a MonoLead, make sure the
number faces in the same direction as the M540
display.
NOTE
When using a 12-lead ECG where the lead wires
are coiled, it is recommended that the 6-wire lead
set is coiled in the same direction as the 4-wire
lead set to prevent artifact. For example, both
lead sets are either coiled towards the patient or
away from the patient.