Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
371
Scio Monitoring
O
2
/N
2
O parameter field
The O
2
/ N
2
O parameter field displays the following
additional values:
– Inspired N
2
O
(inN
2
O) – the level of N
2
O in the
airway during inspiration, taken as the minimum
value within the measurement interval
– End-tidal N
2
O
(etN
2
O) – the highest level of
N
2
O in the airway during the expiratory phase
within the measurement interval
A
O
2
label
B
N
2
O label
C
Abbreviation for "inspiratory"
D
inO
2
(FiO
2
)
value – the level of O
2
in the airway
during inspiration, taken as the minimum value
within the measurement interval
E
inN
2
O value – the level of N
2
O
in the airway
during inspiration, taken as the minimum value
within the measurement interval
F
inO
2
(FiO
2
)
alarm limits (displays crossed
triangle symbol when inO
2
alarms are turned
off)
G
Abbreviation for "expiratory"
H
etO
2
value – the highest level of O
2
in the
airway during the expiratory phase within the
measurement interval
I
etN
2
O value – the highest level of N
2
O in the
airway during the expiratory phase within the
measurement interval
J
etO
2
alarm limits (displays crossed triangle
symbol when etO
2
alarms are turned off)
O
2
_N2O
_pbox.
25
O2
100
20
in
et
30
100
10
N2O
70
69
A
B
C D E F
G
H I J