Troubleshooting
564
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Low
%0 value temporarily
unavail.
O
2
parameter has unknown
accuracy, possibly due to:
– Polluted ambient air
during zeroing.
– Electromagnetic distur-
bances.
– Overheating.
– Ensure clean ambi-
ent air.
– Remove radiating de-
vices (e.g., tele-
phone).
– Check ambient tem-
perature.
– Check water trap and
sample line.
– Change the water
trap or sample line if
necessary.
– Wait for auto zeroing
to complete.
– Power cycle the gas
analyzer.
– Change vaporizer
settings.
– Call DrägerService
®
.
Medium
RRc > #
Respiratory rate has exceed-
ed the limit.
Check ventilation.
Medium
RRc < #
Respiratory rate is below the
limit.
Check ventilation.
Medium
RRc apnea
No breathing or ventilation.
– Start manual ventila-
tion.
– Check ventilation set-
tings.
– Check spontaneous
breathing ability of
the patient
Sample line is not connected. Connect sample line to
breathing circuit
Medium
%0 out of range high
RRc has exceeded the upper
limit of the measuring range
for Scio.
Check ventilation.
Low
Sample line blocked
Sample line or patient-side
filter is occluded.
Check sample line, water
trap, and patient-side
filter.
Alarm Priority Alarm
Cause
Remedy