Getting started
104
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
To discharge a patient
1
Select the left most field on the header bar of
the Cockpit to access the
Demographics
page.
2
Select the
Start
tab (if not already selected).
3
Select the
Discharge
button.
or
1
Select
Start/Standby...
on the main menu bar.
2
Select the
Start
tab (if not already selected).
3
Select the
Start
button. A pop-up window with
the message
Caution discharge will delete
patient data
appears.
4
Select the
Discharge
button in the pop-up
window.
Discharging a patient may take some time
during which the message
Please wait...
ap-
pears on the screen. Once the patient is dis-
charged, the message
Touch Screen to initi-
ate monitoring
appears in the center of the
screen. The message
Patient transferred
ap-
pears in the
Alarm history
page.
Patient categories
Each patient category has specific profiles
associated with it. Profiles are a set of patient and
user settings that have been pre-configured by the
factory or the hospital (for more information, see
"Patient and default profiles" on page 75).
The Cockpit supports the following patient
categories:
If an M540 docks with a different patient category
from the one selected on the Cockpit, the following
happens:
– The Cockpit aligns its patient category to the
M540 patient category setting.
– During the patient category alignment, the
M540 continues to monitor the patient.
– The profile changes to the default profile for the
new patient category and the message
Please
wait ...
is displayed.
– As soon as the Cockpit has switched to the new
patient category, the patient data are automati-
cally transferred to the Cockpit from the M540
that has been monitoring the patient.
NOTE
If configured to appear on the main menu bar, the
Discharge
button is also accessible on the main
menu bar. For more information, see page 452.
Patient
category
Typical Age
Range
Weight
Height
Adult
12 to 140 years
0.1 to 350.0 kg (0.1 to 772.0 lbs) 10 to 250 cm (5 to 100 in)
Pediatric
0 to 16 years
0.1 to 350.0 kg (0.1 to 772.0 lbs) 10 to 250 cm (5 to 100 in)
Neonate
0 to 2 years
1 to 10,000 g (0.1 oz to 351 oz)
10 to 250 cm (5 to 100 in)