Scio Monitoring
370
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
O
2
display
On the Cockpit, the O
2
display consists of:
– O
2
parameter field
– O
2
waveform
The Cockpit supports a configurable parameter
field for display of O
2
that allows one of the
following displays:
– O
2
by itself (default)
– O
2
with N
2
O
O
2
parameter field
The default O
2
parameter field displays the current
values for:
– Inspired O
2
(inO
2
/FiO
2
) – the level of O
2
in the
airway during inspiration phase.
– End-tidal O
2
(etO
2
) – the level of O
2
in the
airway at the end of expiration.
A
O
2
label
B
Abbreviation for "inspiratory"
C
inO
2
(FiO
2
)
value – the level of O
2
in the airway
during inspiration, taken as the minimum value
within the measurement interval
D
inO
2
(FiO
2
)
alarm limits (displays crossed
triangle symbol when inO
2
alarms are turned
off)
E
Abbreviation for "expiratory"
F
etO
2
value – the highest level of O
2
in the
airway during the expiratory phase within the
measurement interval
G
etO
2
alarm limits (displays crossed triangle
symbol when etO
2
alarms are turned off)
NOTE
O
2
monitoring is available only with Scio Four Oxi
and Scio Four Oxi plus.
O
2_pbox
_en.
25
O2
100
20
in
et
50
100
10
A
B C D
E
F G