![Dräger Infinity Acute Care System M540 Скачать руководство пользователя страница 125](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773125.webp)
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
125
Alarms
To deactivate alarm monitoring permanently
1
Select the
symbol next to the
Alarms...
but-
ton on the main menu bar of the Cockpit.
2
Select the
All alarms off
button on the toolbar.
To activate alarm monitoring after deactivating
1
Select the
symbol next to the
Alarms...
but-
ton on the main menu bar of the Cockpit.
2
Select the
All alarms off
button again on the
toolbar.
The Cockpit provides acoustic and optical alarm
signals again when it detects a new alarm
condition.
Configuring the alarm settings for a patient
The following section describes the available alarm
features and settings. You can adjust the alarm
settings for an individual parameter in the
respective parameter-specific setup page. Or, you
can set up the alarm settings of multiple parameters
in one page. When setting alarm limits, make sure
that they are appropriate for the patient’s condition.
Activating/deactivating alarms
Except for the following parameters, you can
activate or deactivate the alarm function for
individual parameters:
– Asystole and ventricular fibrillation (for these
arrhythmia events you cannot deactivate
alarms unless the
ASY/VF alarms
feature is set
to
Follow HR alarm
)
– Cardiac output (C.O.)
– Injectate temperature (Tinj)
– Pulmonary wedge pressure (PWP)
– Paced beats (%PACED)
– Perfusion index (PI)
– Total oxygen content (SpOC) for the Masimo
rainbow SET MCable
– Parameters originating from a device that is
displaying its values on the Cockpit using the
device connectivity option.
NOTE
If configured to appear on the main menu bar, the
All alarms off
button is also accessible on the
main menu bar. For more information, see
page 452.
NOTE
If configured to appear on the main menu bar, the
All alarms off
button is also accessible on the
main menu bar. For more information, see
page 452.