![Dräger Infinity Acute Care System M540 Скачать руководство пользователя страница 328](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773328.webp)
Cardiac output (C.O.)
328
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
The following tables list the computation constants
for Baxter, BD/Ohmeda, and Arrow catheters.
Baxter computation constants
Injectate temperature (Tinj)
sensor connected
sensor disconnected
Catheter
size
Injectate
volume
Tinj = -5 to + 16.0 °C
(23 to 60.9 °F)
Tinj = 16.1 to 25 °C
(61.0 to 80 °F)
Tinj = 20 °C (68 °F)
7F
10 cc
0.561
0.608
0.542
7F
5 cc
0.259
0.301
0.247
7.5F
10 cc
0.574
0.595
0.564
7.5F
5 cc
0.287
0.298
0.257
5F
5 cc
0.285
0.307
0.270
BD/Ohmeda computation constants
Injectate temperature (Tinj)
sensor connected
Catheter
size
Injectate
volume
Tinj = -5 to + 16.0 °C
(23 to 60.9 °F)
Tinj = 16.1 to 25 °C
(61.0 to 80 °F)
7.5F
10 cc
0.579
0.628
7.5F
5 cc
0.281
0.309
7.5F
3 cc
0.160
0.181
7F
10 cc
0.579
0.628
7F
5 cc
0.281
0.309
7F
3 cc
0.160
0.181
5F
5 cc
0.291
0.316
5F
3 cc
0.170
0.188