Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
531
Troubleshooting
None
SpO2 Low
Perfusion
Low perfusion
Parameter values
The signal is too
small.
– Check the
patient and
treat if
necessary.
– Move the
sensor to a site
that is more
adequately
perfused.
!
SpO2 MCable
unplugged
2)
MCable
unplugged
Parameter values
are replaced by
***. The parameter
values are
replaced by blanks
for PI or SpOC if
using Masimo
rainbow SET
The SpO2 MCable is
disconnected from
the M540.
Check connections
to M540.
None
(Masimo
rainbow
SET
MCable
only)
SpO2 only mode
SpO2 only mode
Parameter values
for SpO2, PLS, PI,
and PVI.
Parameter values
for SpHb (SpHbv),
SpOC, SpMet,
SpCO are
replaced by ***
The device cannot
calibrate the Masimo
rainbow SET
parameters and is
attempting to display
the standard Masimo
parameters.
Remove and
reapply the sensor.
If the problem
persists, contact
Dräger-authorized
service personnel.
!
(Any
Masimo
MCable)
SpO2 replace cable
next pt.
Replace cable
next pt.
Parameter values
Cable expired.
Replace the cable.
!
(Any
Masimo
MCable)
SpO2 replace sensor
next pt.
Replace sensor
next pt.
Parameter values
– SpO
2
sensor
expired.
– Adhesive sensor
expired.
Replace the
sensor.
2)
After discharging a patient or starting the device, the alarm becomes active only after a numeric value
has been received for that parameter.
Priority
Message
Parameter field
Problem
Solution