Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
475
System configuration
Biomed setup
–
name service settings
The following table lists the available settings of the
Name service
page. To access this page, see
page 470. After making the desired changes, select
the
Apply
button which causes a brief loss of
communication with a docked M540.
Selection
Available settings
Description
Monitoring unit ID
1 to 255 (increments of 1)
Allows you to assign the Cockpit to a
monitoring unit by entering an ID using the
keypad
symbol.
Monitoring unit
label
Up to seven alphanumeric
characters
Allows you to enter the corresponding label for
the network and recordings using the
keyboard symbol.
When entering the
Bed label
, do not use the
word’IACS’.The word’IACS’ is used for the
Name service
setting. If the
Name service
setting and any bed labels match, the IACS
bed label does not appear on the ICS.
Care unit label
Bed label
Hospital name
Enable Central
Station
–
On
(default)
– Off
If this feature is activated and an ICS is
connected to the network, the message
Not
Monitored By Central
appears in the Cockpit
header bar if the Cockpit is not assigned to an
ICS. For more information, (see page 86).
If this feature is deactivated, an ICS is
connected to the network and the Cockpit is
not assigned to an ICS, the message
Not
Monitored By Central
does not appear in the
Cockpit header bar.
Enable Remote
Control
–
On
(default)
– Off
If this feature is activated, the Cockpit allows
other Infinity monitors and the ICS to view its
data and perform simple functions, such as
requesting a recording or pausing an alarm.
Enable Remote
Silence
–
On
(default)
– Off
If this feature is activated, the Cockpit allows
alarms to be silenced from network devices.