Troubleshooting
550
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Sample line
Priority
Message
Cause
Remedy
!
Sample line is being
cleared
A sample line blockage
occurred and the
Microstream MCable is
attempting to clear the
sample line.
Informational message – no
action required.
!
Sample line blocked
The sample line is blocked
during the purging process.
Replace the sample line.
!
Sample line
disconnected
The sample line is
disconnected from the
Microstream MCable.
Securely connect the sample
line to the Microstream
MCable
.