Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
113
Alarms
Optical alarm indicators on the Cockpit
A
Alarm message field in the blue header bar
B
Alarming parameter field
C
Alarm bar
Optical alarm indicators on the M540
Alarm bar
The alarm bar on the Cockpit and the M540
optically announces high and medium-priority
alarm conditions (see page 109). The color of the
alarm bar always reflects the priority of the alarm
condition. It may change from yellow to red and vice
versa depending on the latest alarm condition. The
alarm bar appears in solid color for any
unacknowledged single notification alarm.
However, the alarm bar is inactive when:
– Only low-priority alarm conditions exist
– The alarm bar is deactivated (see page 460)
– Cardiac bypass or Privacy modes are activated
(see page 119)
– Alarm monitoring is deactivated (see page 125)
Header bar
The header bar displays the alarm message on the
background color corresponding to the alarm
priority. During multiple alarm conditions, the
optical alarm signals always reflect the condition
corresponding to the highest alarm priority. The
header bar accommodates up to two messages
simultaneously (see page 110).
016
B
C
A
30
9
B
A
C
NOTE
The color of the alarm bar always corresponds to
the highest priority alarm condition for all active or
audio pause alarms in audio pause.