Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
47
Operating concept
Docking to a different Cockpit
If the user undocks an M540 from a Cockpit and
later docks the M540 to a different Cockpit, the
original data are automatically retrieved over the
network by the new Cockpit. The new Cockpit then
automatically merges this data with any data the
M540 collected while on patient transport. The
original Cockpit automatically discharges the
patient once all patient data are transferred.
When moving the M540 between Cockpits in
different monitoring units, the time stamps may
differ occasionally between the Cockpit and the
M540.
If not all patient data was transferred, the message
Transfer of Data Incomplete
appears in the
header bar of the new Cockpit. In this case, the
original Cockpit does not discharge the patient.
Undocking the M540
When the user undocks the M540, the following
happens:
– The message
Disconnected from M540
appears in the center of the Cockpit screen.
– When the M540 is not in wireless mode, a
message appears at the ICS that the bed is
disconnected. When the wireless option for an
M540 is activated and configured properly, the
ICS displays the wireless symbol.
– Data are no longer trended at the Cockpit.
– With each docking or undocking of an M540,
there is a short transition period. One minute of
trend data collected during this transition period
may not be displayed at an ICS equipped with
software version VG1. However,
this
trend data
can be reviewed at the Cockpit.
– Several buttons remain active on the main
menu bar of the Cockpit:
–
Alarms...
for accessing the alarm history.
–
Trends/ Data...
for accessing the trend
data.
–
Start/Standby...
for accessing the
Start
tab
from where the user can initiate a patient
discharge.
– The current patient data from the Cockpit are no
longer available to the Infinity network.
However, when the wireless option for an M540
is activated and configured properly, it
continues to make the data available to the
Infinity network.
– Parameter values acquired using the device
connectivity option are no longer available to
the Infinity network.
CAUTION
Before connecting the M540 to a different
Cockpit, make sure that the units of measure align
between the two devices. Differing units of
measure could result in loss of data or a patient
discharge.