Alarms
154
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
IBP D neonate
Increment:
1 mmHg
or 0.1 kPa
(GP1 D to
GP4 D,
LV D,
RV D)
On
(ART D,
PA D)
Upper: –24 to +300 mmHg
–3.2 to +40.0 kPa
Lower: –25 to +299 mmHg
–3.3 to +39.9 kPa
– 80 mmHg
(10.7 kPa)
for GP1 D to
GP4 D, PA D
– 25 mmHg
(3.3 kPa) for
LV D
– 13 mmHg
(1.7 kPa) for
PA D, RV D
– 35 mmHg
(4.7 kPa)
for GP1 D
to
GP4 D, PA
D
– 2 mmHg
(0.3 kPa)
for PA D,
LV D,
RV D
Off
IBP M adult
Increment:
1 mmHg
or 0.1 kPa
On
Upper: –24 to +300 mmHg
–3.2 to +40.0 kPa
Lower: –25 to +299 mmHg
–3.3 to +39.9 kPa
– 125 mmHg
(16.7 kPa)
for GP1 M to
GP4 M,
ART M
– 80 mmHg
(10.7 kPa)
for LV M
– 20 mmHg
(2.7 kPa) for
LA, ICP, CVP
– 17 mmHg
(2.3 kPa) for
PA M, RV M
– 12 mmHg
(1.6 kPa) for
RA
– 60 mmHg
(8.0 kPa)
for GP1 M
to
GP4 M,
ART M
– 40 mmHg
(5.3 kPa)
for LV M
– 7 mmHg
(0.9 kPa)
for PA M,
RV M
– 2 mmHg
(0.3 kPa)
for RA, ICP
– 0 mmHg
(0.0 kPa)
for LA,
CVP
Off
Parameter Alarm
default
status
Alarm limit range
Upper limit
defaults
Lower limit
defaults
Archive
default
setting