Alarms
160
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Arrhythmia ranges and defaults
Parameter
Alarm priority
default
Rate (default)
Count (default)
Alarm archive
factory default
ASY
High
Not adjustable
Not adjustable
Str/Rec
VF
High
Not adjustable
Not adjustable
Str/Rec
VTACH
High
≥
100 to 200 (
≥
120)
Increments of 10
≥
5 to 15 (
≥
10)
Increments of 1
Str/Rec
ARTF
Off
Not adjustable
Not adjustable
Off
RUN
Medium
not adjustable (Rate =
VTACH)
3 to VT count – 1 (3 to
9) changes based on
VTACH
Str/Rec
AIVR
Medium
Not adjustable =
VTACH rate – 1 (
≤
119
)
Not adjustable (
≥
3)
Off
SVT
Medium
≥
120 to 200 (
≥
150)
Increments of 10
≥
3 to 10 (
≥
3)
Increments of 1
Str/Rec
CPT
Low
Not adjustable
Not adjustable
Str/Rec
BGM
Low
Not adjustable
Not adjustable
Str/Rec
TACH
Off
≥
100 to 200 (
≥
130)
Increments of 10
≥
5 to 15 (
≥
8)
Increments of 1
Off
BRADY
Off
≤
30 to 105
(adult
≤
50; pediatric
≤
60)
Increment of 5
Not adjustable
(
≥
8)
Off
Pause
Off
1 to 3.5 (2.5)
Increments of 0.5
Not adjustable
Off