Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
591
Technical data
Infinity MCable – Nurse call
Physical attributes
Connections
Connects to the PS250 / P2500
Connection via cable 8417370 only
Cable signals during non-alarm state
Cable 1 (NO normally open): white
Cable 2 (COM common): brown
Cable 3 (NC normally closed): green
Mode of operation
Continuous
Power requirements
Input voltage
24 V DC maximum
Input current
1 A DC maximum
Switching capacity
15 W maximum
Protection against electric shock
Three contacts from the open cable have a
galvanic isolation of 1.5 k V AC
Environmental requirements
Operation
Temperature
5 to 55 °C (41 to 131 °F)
Relative humidity
5 to 95%, non-condensing
Ambient pressure
375 to 825 mmHg (50 to 110 kPa)
Storage
Temperature range
–20 to +60 °C (–4 °F to +140 °F)
Relative humidity
5 to 95%, non-condensing
Ambient pressure
375 to 825 mmHg (50 to 110 kPa)