System configuration
468
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Recorder setup
The following table lists the available settings of the
Recorder setup
page.
Selection
Available settings
Description
Delay
6, 10 (default), 15 s
Determines the amount of delay (pre-event)
data included in a timed recording. Delay data
refers to data that originated
before
the
recording was initiated. A marker on the strip
recording marks where the delay data ends
and the real-time data starts.
Duration
6, 10, 15, 20 s (default)
Determines the length of a timed recording.
Speed
6.25, 12.50, 25.00 (default),
50.00 mm/s
Determines the recording speed.
Waveform
Selection
–
Auto
(default)
– Manual
–
Auto
– the top two displayed waveforms
are automatically selected for recordings. If
no waveforms are displayed, no recording
is generated.
–
Manual
– the two selected waveforms
under
Waveform 1
and
Waveform 2
are
printed.
Waveform 1
Selected parameter under
Waveform Selection
setting
(factory default is ECG Lead II)
Assigns the selected waveform to the top
channel on R50N recordings, provided the
Waveform Selection
is set to
Manual.
Waveform 2
Selected parameter under
Waveform Selection
setting
(factory default is ECG lead V)
Assigns the selected waveform to the bottom
channel on R50N recordings, provided
Waveform Selection
is set to
Manual.
Alarm Waveform
–
On
(default)
– Off
When this function is activated, the waveform
of an alarming parameter of medium or high
priority is printed in the second recording
channel provided the archive function is
activated.