Technical data
590
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
LEDs
– Green LED (device is connected to AC power).
– Yellow LED (briefly during startup)
– The battery/charging failure indicator lights up yellow
under the following circumstances:
– Battery failure (battery depleted, battery not connect-
ed, battery fault), or
– Battery charging error, or battery temperature issue
The Cockpit LEDs on the front light up for a few seconds
when the Infinity P2500 is connected to AC mains after it has
been disconnected for an interval to indicate that the LED is
working properly.
Automatic shutdown
– Overload limit (the device shuts down)
– Power overload: exceed 285 W (at system connector)
– Current overload: exceed 14 A (short circuit protected)
– Output voltage: exceed 32V
– Temperature:
– internal temperature exceeds 75°C (167 °F)
– battery temperature exceeds 50 °C (122°F) when
battery is in charging status after 1 minute.
Electrical protection
Class 1
Inrush current
37 A