Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
269
SpO2 and Pulse CO-Ox monitoring with Masimo SET MCable
Password-protected Masimo rainbow SET setup functions
Additional Masimo rainbow SET setup functions
take place in the
Setup
page which is protected by
a clinical password. To access this setup page, see
page 263.
Selection
Available settings
Description
SpHb Cal
–
Arterial
(default)
– Venous
Selects the blood sampling
source which is used to
calculate the SpHb value.
The SpHb value changes to
SpHbv when the
SpHb Cal
setting
Venous
is selected.
PVI averaging
time
– Short
–
Long
(default)
Determines how responsive the
monitor is to rapid physiological
changes while tracking pleth
variability index.
A longer averaging time
provides a more accurate
result. However, in clinical
situations where rapid
physiological changes have to
be monitored, use a shorter
averaging time.