Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
527
Troubleshooting
!!
(Any
SpO2
MCable)
PLS out of range low
The parameter
value is replaced
by - - -
The parameter value
is below the
measurement range
of the monitor.
– Check the
patient and
treat if
necessary.
– Change the
alarm limits.
PLS out of range
high
The parameter
value is replaced
by +++
The parameter value
is above the
measurement range
of the monitor.
!!
(Masimo
rainbow
SET only)
PVI
> (alarm limit)
SpHb
> (alarm limit)
SpHbv
> (alarm limit)
SpMet
> (alarm limit)
SpOC
> (alarm limit)
Parameter value
The parameter value
is above the upper
alarm limits.
– Check the
patient and
treat if
necessary.
– Change the
alarm limits.
!!
(Any
SpO2
MCable)
SpO2
> (alarm limit)
PLS
> (alarm limit)
!!
(Masimo
rainbow
SET only)
SpHb
< (alarm limit)
SpHbv
< (alarm limit)
PVI
< (alarm limit)
SpOC
< (alarm limit)
SpMet
< (alarm limit)
Parameter value
The parameter value
is above/below the set
upper/lower alarm
limits.
The priority changes
to high (!!!) if the SpO2
value falls more than
10% below the lower
limit. This does not
occur when using
SatSeconds alarm
time with the Nellcor
OxiMax MCable.
– Check the
patient and
treat if
necessary.
– Change the
alarm limits.
(Any
SpO2
MCable)
SpO2
< (alarm limit)
PLS
< (alarm limit)
Priority
Message
Parameter field
Problem
Solution