![Dräger Infinity Acute Care System M540 Скачать руководство пользователя страница 131](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773131.webp)
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
131
Alarms
Configuring the alarm message behavior
The
Config.
tab allows you to set up the acoustic
and optical alarm signal reporting behavior in
response to certain SpO2 sensor and messages
related to leads-off conditions.
To access the
Config.
tab
1
Select the
Alarms...
button on the main menu
bar.
2
Select the
Limits
tab (if not already selected).
3
Select the
Config.
tab
4
Refer to the following table for available
settings. The selected alarm priority affects how
the alarm event is reported optically and
acoustically (see page 114).
The alarm setting:
– is supported on the ICS
– is supported in any remote view
– is stored in the patient profile
Regardless of the selected setting, the acoustic
alarm signal can be audio paused but resumes if
the condition persists beyond the two minute audio
pause time. The message appears in the header
bar of the Cockpit until the condition disappears or
you acknowledge.
Message
Available settings
Description
SpO2 sensor off
1)
SpO2 check sensor
2)
Alarm settings
– High
– Medium
–
Low
(default for
SpO2 check sensor
)
– One-shot
–
(off) – default for
SpO2 sensor off
Assigns an alarm priority to the sensor alarm or
deactivates the sensor alarm.
–
High:
The event is treated as a high-priority
alarm.
–
Medium
: The event is treated as a medium-
priority alarm.
–
Low
: The event is treated as a persistent low-
priority alarm.
–
One-shot
: The event is treated as a low priority,
single notification alarm.
–
: No optical or acoustic alarm signals are
triggered; however, if the sensor is no longer
attached to the patient, a corresponding
message appears in the SpO
2
parameter field.
1)
Message originating from a Masimo rainbow SET or a Masimo SET MCable.
2)
Message originating from a Nellcor MCable.