Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
581
Reprocessing
Reprocessing procedures
Validated reprocessing procedures
The effectiveness of the listed reprocessing
procedures has been validated by independent
laboratories that are certified to the standard ISO
17025.
At the time of validation, the following reprocessing
procedures showed good material compatibility
and effectiveness:
Disinfection with cleaning
1
Remove soiling immediately. Use a cloth
dampened with cleaning agent to remove
soiling.
2
Disinfect the surface.
3
After the product has been exposed to the
disinfectant for the specified contact time,
remove residual disinfectant.
4
Wipe with a cloth dampened with water
(preferably drinking-water quality). Allow the
product to dry.
5
Check the product for visible soiling. Repeat
steps one through four if necessary.
6
Check the product for visible damage and
replace if necessary.
Procedure
Agent
Manufacturer
Concentration
Contact time
Temperature
Disinfection
with
cleaning
Buraton
10 F
Schülke &
Mayr
1%
30 min
N/A
Dismozon
pur
BODE
Chemie
1.5%
15 min
N/A
WARNING
Risk of electric shock and device malfunction
Penetrating liquid may cause the following:
- Damage to the device
- Electric shock when switching on the device
- Device malfunctions
Ensure that no liquid penetrates the device.